Brief introduction:
servo-controlled by micro-computer; quite operation windows can reduce the disturbance to the baby during nursing; touch switch panel with LED digital display, set-temp, real-temp are displayed separately; multiply self-checking alarms with audible and visible; heating power rate indicates 0-100%.

Standard configure:
Hood, infant bed, temperature controller, water tank, air filter, blue light radiant box, I.V. Pole, stand.

Technical specification:
Power supply : AC220V,50HZ
Power input: 600VA
Air Temperature control range:25℃~37℃
Changing of temperature: ≤0.5℃
Uniformity of temperature: ≤0.8℃
Temperature rising time: ≤45min
Noise level: ≤55dB(A)

Alarming:
audible and visible alarm indication, and it can indicate the real temperature and cut off the power supply of heater.
Over-temp alarm: when the temperature control ≤38℃;
Deviation alarm:±3℃;
Sensor alarm: when the sensor is disconnected, open circuit, short circuit or put on the wrong place;
Fan alarm: when the fan is blocked, speed ≤1000rpm or stop working;
Power failure alarm: when there is no power or the power cord is disconnected.
System alarm: when there is E2ROM malfunction inside of temperature controller.

Package:
Gross Weight:72Kg
Measurement:1085mm×640mm×1065mm


 

 
YP-100A
INFANT INCUBATOR
State Food and Drug Administration 2005 No.: 3540651
CMDC TUV ISO9001 ISO13485 CE0123
 
 
 
Brief introduction:
servo-controlled by micro-computer; quite operation windows can reduce the disturbance to the baby during nursing; touch switch panel with LED digital display, set-temp, real-temp are displayed separately; multiply self-checking alarms with audible and visible; heating power rate indicates 0-100%; the infant bed can be pulled out from the side door

Standard configure:
Hood, infant bed, temperature controller, water tank, air filter, I.V. Pole, stand.

Technical specification:

Power supply : AC220V,50HZ
Power input: 600VA
Air Temperature control range:25℃~37℃
Changing of temperature: ≤0.5℃
Uniformity of temperature: ≤0.8℃
Temperature rising time: ≤45min
Noise level: ≤55dB(A)

Alarming:
audible and visible alarm indication, and it can indicate the real temperature and cut off the power supply of heater.
Over-temp alarm: when the temperature control ≤38℃;
Deviation alarm:±3℃;
Sensor alarm: when the sensor is disconnected, open circuit, short circuit or put on the wrong place;
Fan alarm: when the fan is blocked, speed ≤1000rpm or stop working;
Power failure alarm: when there is no power or the power cord is disconnected.
System alarm: when there is E2ROM malfunction inside of temperature controller.

Package:
Gross Weight:72Kg
Measurement:1085mm×640mm×1065mm

 

 
 
 
YP-100B
INFANT INCUBATOR
State Food and Drug Administration 2005 No.: 3540651
CMDC TUV ISO9001 ISO13485 CE0123
 
 
 
Brief introduction:
servo-controlled by micro-computer; quite operation windows can reduce the disturbance to the baby during nursing; touch switch panel with LED digital display, set-temp, real-temp are displayed separately; multiply self-checking alarms with audible and visible; heating power rate indicates 0-100%; blue light radiant box can give the baby jaundice phototherapy treatment when keeping warm on baby with blue light timer.

Standard configure:

Hood, infant bed, temperature controller, water tank, air filter, blue light radiant box, I.V. Pole, stand.

Technical specification:
Power supply : AC220V,50HZ
Power input: ≤600VA
Temperature control range:25℃~37℃
Changing of temperature: ≤0.5℃
Uniformity of temperature: ≤0.8℃
Temperature rising time: < 45 min
Noise level: ≤55dB(A)
Average radiation of blue light: ≥500μW/cm2
Timing range of blue light timer:0~9999 hour 59 minutes
Accumulative error of blue light timing:≤0.002%

Alarming:
audible and visible alarm indication, and it can indicate the real temperature and cut off the power supply of heater.
Over-temp alarm: when the air temperature control ≤38℃;
Deviation alarm: air temperature control ±3℃;
Sensor alarm: when the sensor is disconnected, open circuit, short circuit or put on the wrong place;
Fan alarm: when the fan is blocked, speed ≤1000rpm or stop working;
Power failure alarm: when there is no power or the power cord is disconnected.
System alarm: when there is E2ROM malfunction inside of temperature controller.

Package:
Gross Weight:83Kg
Measurement:1085mm×640mm×1110mm

 

 
 
YP-100AB
INFANT INCUBATOR
State Food and Drug Administration 2005 No.: 3540651
CMDC TUV ISO9001 ISO13485 CE0123
 
 
 
Brief introduction:
servo-controlled by micro-computer; quite operation windows can reduce the disturbance to the baby during nursing; touch switch panel with LED digital display, set-temp, real-temp are displayed separately; multiply self-checking alarms with audible and visible; heating power rate indicates 0-100%; the infant bed can be pulled out from the side door; blue light radiant box can give the baby jaundice phototherapy treatment when keeping warm on baby with blue light timer.

Standard configure:
Hood, infant bed, temperature controller, water tank, air filter, blue light radiant box, I.V. Pole, cabinet.

Technical specification:
Power supply : AC220V,50HZ
Power input: ≤600VA
Temperature control range:25℃~37℃
Changing of temperature: ≤0.5℃
Uniformity of temperature: ≤0.8℃
Temperature rising time: < 45 min
Noise level: ≤55dB(A)
Average radiation of blue light: ≥500μW/cm2
Timing range of blue light timer:0~9999 hour 59 minutes
Accumulative error of blue light timing:≤0.002%

Alarming:
audible and visible alarm indication, and it can indicate the real temperature and cut off the power supply of heater.
Over-temp alarm: when the air temperature control ≤38℃;
Deviation alarm: air temperature control ±3℃;
Sensor alarm: when the sensor is disconnected, open circuit, short circuit or put on the wrong place;
Fan alarm: when the fan is blocked, speed ≤1000rpm or stop working;
Power failure alarm: when there is no power or the power cord is disconnected.
System alarm: when there is E2ROM malfunction inside of temperature controller.

Package:
Gross Weight:83Kg
Measurement:1085mm×640mm×1110mm
 

 
 
HKN-93A
INFANT RADIANT WARMER
State Food and Drug Administration 2005 No.: 3540553
CMDC TUV ISO9001 ISO13485 CE0123
 
 
 
Brief introduction:
servo-controlled by micro-computer with option of automatic control or manual control the temperature; skin temperature sensor and bed temperature sensor; touch switch panel, setting temperature, real temperature, and APGAR timer can display separately, heating power rate indicates 0-100%; multi-function self-checking alarms; horizon angle of radiant head can be adjusted in two ways, there are 2pcs of radiant boxes; inclination of bassinet is adjustable; x-ray cassette is under the bed; 2pcs of oxygen cylinders and oxygen infusion device; RS232 connector, data output.

Standard configure:
radiation head, blue light radiant box (2pcs), temperature controller, skin temperature sensor, bed temperature sensor, infant bed, I.V. Pole, tray, x-ray cassette, oxygen cylinders, oxygen infusion device, low-pressure aspirator, resuscitation bag, infant head fixing frame, Stand.

Technical specification:
Power supply : AC220V 50HZ
Power input: 950VA
Temperature control range:25℃~37℃
Sensor precision: ≤0.3℃
Uniformity of bed temperature: ≤2.0℃
Adjustment angle for the radiation head :0°,30°,60
inclination of infant bed: ±15°
APGAR Timer: it has sound indication in 1min, 5min, 10min.

Alarming:
with audible and visible alarm, and it can indicate the real temperature and cut off the power supply of heater;
Over-temp alarm: when the skin temperature >39.5℃;
Deviation alarm: indication temperature deviates away the setting temperature ±1℃;
Sensor alarm: when the sensor is disconnected, open circuit, short circuit or put on the outside of bed;
Power failure alarm: when start up the unit, but there is no power and the power cord is disconnected.
System alarm: when there is system malfunction inside of temperature controller.

Package:
Gross Weight:150Kg
Measurement:1270mm×810mm×1110mm
 

 
XHZ
NEONATE BILIRUBIN PHOTOTHERAPY EQUIPMENT
ZJFDA(Z) Z 2007 No.: 2260333
CMDC TUV ISO9001 ISO13485
 
 
 
Brief introduction:
Temperature servo-controlled by micro-computer, and this unit Adopts the far-infrared ceramic heating plate as heating resource, has the function of over-temp alarm and sensor alarm; blue light in double way, there are total 11 pieces of blue lamp with 20W for each; touched operation panel, LCD can display clock and temperature with blue light timer; and inside of box, there is heat-dissipating fan; the infant bed is made of whole transparent acryl plastic, and the panels around the bed can be opened, and the infant bed can be pulled out from the side door.

Standard configure:
Temperature controller, Up Cabinet, Cabinet, infant bed, I.V.Pole.

Technical specification:
Power supply : AC220V 50HZ
Power input: 800VA
Temperature control range: 25℃~37℃
The margin between the indication temperature and the control temperature: ≤0.7℃
Uniformity of the surface temperature on bed: ≤3.5℃
Radiation of Blue light for upper radiant box:≥850μW/cm2
Radiation of Blue light for lower radiant box:≥700μW/cm2
Timer range of blue light:0~9999 hours ( Reset )
Over temperature alarm: ≤39.5℃, give the visible and audible alarm
Sensor Alarm: The leads of sensor plug is open circuit or close circuit.

Package:
Gross Weight:Main body 70Kg / Cabinet 50Kg
Measurement:Main body 1000mm×700mm×790mm
                             Cabinet 910mm×600mm×990mm
 
 
 
XHZ-90
NEONATE BILIRUBIN PHOTOTHERAPY EQUIPMENT
ZJFDA(Z)Z 2005 No.: 2260262
CMDC TUV ISO9001 ISO13485
 
 
 
Brief introduction:
The pillar is made of stainless steel, has the variable height adjustment mechanism and swivel castors with brake; can be equipped with the infant incubator or the infant bed; and the angle of radiant box can be adjusted, inside of this box, there are 5 pieces of blue lamp with 20W for each; LCD display Timer can record the treatment time or the using time for the lamp accurately.

Standard configure:
Radiant box, pillar, lift pole, stand.

Technical specification:
Power supply : AC220V 50HZ
Power input: ≤100VA
Blue light wavelength:420nm~470nm
Average radiation of blue light:≥500μW/cm2
Adjustment height for the radiation box:1350mm~1650mm
Timer range of blue light:0~9999 hours ( Reset )

Package:
Gross Weight:23Kg
Measurement:955mm×655mm×205mm