Brief
introduction:
servo-controlled by
micro-computer with option of
automatic control or manual
control the temperature; skin
temperature sensor and bed
temperature sensor; touch switch
panel, setting temperature, real
temperature, and APGAR timer can
display separately, heating
power rate indicates 0-100%;
multi-function self-checking
alarms; horizon angle of radiant
head can be adjusted in two
ways, there are 2pcs of radiant
boxes; inclination of bassinet
is adjustable; x-ray cassette is
under the bed; 2pcs of oxygen
cylinders and oxygen infusion
device; RS232 connector, data
output.
Standard configure:
radiation head, blue light
radiant box (2pcs), temperature
controller, skin temperature
sensor, bed temperature sensor,
infant bed, I.V. Pole, tray,
x-ray cassette, oxygen
cylinders, oxygen infusion
device, low-pressure aspirator,
resuscitation bag, infant head
fixing frame, Stand.
Technical
specification:
Power supply : AC220V 50HZ
Power input: 950VA
Temperature control
range:25℃~37℃
Sensor precision: ≤0.3℃
Uniformity of bed temperature:
≤2.0℃
Adjustment angle for the
radiation head :0°,30°,60
inclination of infant bed: ±15°
APGAR Timer: it has sound
indication in 1min, 5min, 10min.
Alarming:
with audible and visible alarm,
and it can indicate the real
temperature and cut off the
power supply of heater;
Over-temp alarm: when the skin
temperature >39.5℃;
Deviation alarm: indication
temperature deviates away the
setting temperature ±1℃;
Sensor alarm: when the sensor is
disconnected, open circuit,
short circuit or put on the
outside of bed;
Power failure alarm: when start
up the unit, but there is no
power and the power cord is
disconnected.
System alarm: when there is
system malfunction inside of
temperature controller.
Package:
Gross Weight:150Kg
Measurement:1270mm×810mm×1110mm
|